Development of an AMR-ACB array for gastrointestinal motility studies.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 22996723.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The association between anisotropic magnetoresistive (AMR) sensor and AC biosusceptometry (ACB) to evaluate gastrointestinal motility is presented. The AMR-ACB system was successfully characterized in a bench-top study, and in vivo results were compared with those obtained by means of simultaneous manometry. Both AMR-ACB and manometry techniques presented high temporal cross correlation between the two periodicals signals . The contraction frequencies using AMR-ACB were 73.9 ± 7.6 mHz and using manometry were 73.8 ± 7.9 mHz during the baseline . The amplitude of contraction using AMR-ACB was 396 ± 108 μT·s and using manometry were 540 ± 198 mmHg·s during the baseline. The amplitudes of signals for AMR-ACB and manometric recordings were similarly increased to 86.4% and 89.3% by neostigmine, and also decreased to 27.2% and 21.4% by hyoscine butylbromide in all animals, respectively. The AMR-ACB array is nonexpensive, portable, and has high-spatiotemporal resolution to provide helpful information about gastrointestinal tract.
Medical subject headings
- Gastrointestinal Motility
- Monitoring, Physiologic
- Signal Processing, Computer-Assisted