Measurement reproducibility of perfusion fraction and pseudodiffusion coefficient derived by intravoxel incoherent motion diffusion-weighted MR imaging in normal liver and metastases.

Andreou, A; Koh, D M; Collins, D J; Blackledge, M; Wallace, T; Leach, M O; Orton, M R · Eur Radiol · 2013

retrospective_cohort · Level III

Where this comes from

Abstract

To determine the measurement reproducibility of perfusion fraction f, pseudodiffusion coefficient D and diffusion coefficient D in colorectal liver metastases and normal liver. Fourteen patients with known colorectal liver metastases were examined twice using respiratory-triggered echo-planar DW-MRI with eight b values (0 to 900 s/mm(2)) 1 h apart. Regions of interests were drawn around target metastasis and normal liver in each patient to derive ADC (all b values), ADC(high) (b values ≥ 100 s/mm(2)) and intravoxel incoherent motion (IVIM) parameters f, D and D by least squares data fitting. Short-term measurement reproducibility of median ADC, ADC(high), f, D and D values were derived from Bland-Altman analysis. The measurement reproducibility for ADC, ADC(high) and D was worst in colorectal liver metastases (-21 % to +25 %) compared with liver parenchyma (-6 % to +8 %). Poor measurement reproducibility was observed for the perfusion-sensitive parameters of f (-75 % to +241 %) and D (-89 % to +2,120 %) in metastases, and to a lesser extent the f (-24 % to +25 %) and D (-31 % to +59 %) of liver. Estimates of f and D derived from the widely used least squares IVIM fitting showed poor measurement reproducibility. Efforts should be made to improve the measurement reproducibility of perfusion-sensitive IVIM parameters.