Catastrophic wear in total hip arthroplasty. Complete wear-through of a metal-backed acetabular component: case report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 23102424.
- Also identified by DOI 10.1016/j.ocl.2012.08.003.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A unique case of catastrophic wear of a total hip arthroplasty is presented. This brief report demonstrates the presentation of catastrophic wear of a modular acetabular component with protrusion of the femoral head through a modular metal-backed titanium cup. During the revision surgery, the polyethylene was found to have disengaged from the cup, and the cup was pistoning on the femoral neck. Significant metallosis and bone loss was found. The presence of severe metallosis mimicked bone on the preoperative radiograph and obscured the severity of the osteolysis seen on imaging.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Prosthesis Failure