Left atrial pressure controller design for an artificial heart.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2312141.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper describes the application of a dynamic compensator technique to the left atrial controller design for use with a portable artificial heart drive system. The compensator is designed using a method in the field of multivariable control. This controller design is based on the physical models of the actuator and blood pump system. The analysis shows that there exists a minimal compensator with a dimension of one. The computer simulation demonstrates the acceptable, robust control performance of the left atrial pressure for a relatively small parameter variation of the vascular system model when all the poles of the closed-loop system are assigned to appropriate values.
Medical subject headings
- Computer Simulation
- Heart, Artificial
- Models, Cardiovascular