Debonding of porous coating of a threaded acetabular component: retrieval analysis.

Łapaj, Łukasz; Markuszewski, Jacek; Rybak, Tomasz; Wierusz-Kozłowska, Małgorzata · J Mech Behav Biomed Mater · 2013

case_report · Level V

Where this comes from

Abstract

This report presents a case of debonding of plasma sprayed porous titanium coating from a threaded acetabular component which caused aseptic loosening of the implant. Weight bearing after delamination caused abrasive damage of the acetabular shell, and particles of the coating embedded in the acetabular liner. Microscopic examination of periprosthetic tissues showed presence of metal particles and macrophage infiltration. Despite microscopic examination of the retrieved component the cause of debonding remains unclear.