Reproducibility of structural, resting-state BOLD and DTI data between identical scanners.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 23133518.
- Also identified by DOI 10.1371/journal.pone.0047684 and PMC identifier 3485040.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Increasingly, clinical trials based on brain imaging are adopting multiple sites/centers to increase their subject pool and to expedite the studies, and more longitudinal studies are using multiple imaging methods to assess structural and functional changes. Careful investigation of the test-retest reliability and image quality of inter- or intra- scanner neuroimaging measurements are critical in the design, statistical analysis and interpretation of results. We propose a framework and specific metrics to quantify the reproducibility and image quality for neuroimaging studies (structural, BOLD and Diffusion Tensor Imaging) collected across identical scanners and following a major hardware repair (gradient coil replacement). We achieved consistent measures for the proposed metrics: structural (mean volume in specific regions and stretch factor), functional (temporal Signal-to-Noise ratio), diffusion (mean Fractional Anisotropy and Mean Diffusivity in multiple regions). The proposed frame work of imaging metrics should be used to perform daily quality assurance testing and incorporated into multi-center studies.
Medical subject headings
- Diffusion Tensor Imaging