In vivo roughening of retrieved total knee arthroplasty femoral components.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 23142273.
- Also identified by DOI 10.1016/j.knee.2012.09.008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Joint registry data highlights the higher rates of cumulative revision for younger patients undergoing TKR. One of the reasons associated with this higher revision rate may be due to the wear of the UHMWPE leading to loosening. Alternate bearing surfaces have been developed to address this problem; however, roughening of the metal bearing surface has not been demonstrated in vivo. We recorded roughness measurements of retrieved femoral components. Average lateral condyle roughness was 0.032 μm, compared to control values of 0.020 μm, p=0.002; average medial condyle roughness was 0.028 μm, compared to a control value of 0.019, p<0.001. There was a small but statistically significant roughening of femoral components in vivo compared to controls. This may have important implications for aseptic loosening of knee arthroplasty components and the decision to use scratch resistant components.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Device Removal
- Knee Prosthesis
- Prosthesis Failure
Anatomy
- femur
- knee