Brachial artery waveforms for automatic blood pressure measurement.
Level II
Where this comes from
- Record sourced from PubMed, PMID 23149078.
- Also identified by DOI 10.1016/j.jbiomech.2012.10.021.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Theoretically the auscultatory method using Korotkoff sounds is more related to the maximum artery closure status, while the oscillometric method is more related to the overall artery closure status under the cuff. Therefore, the latter is less accurate than the former. This work introduces a new method, which is more accurate than the oscillometric method and suitable for automatic devices. To monitor the maximum artery closure status, a piezoelectric film sensor is attached to the skin just above the brachial artery and under the central section of the cuff where maximum cuff pressure is transferred to the arm. Using the waveform features obtained by this sensor, measurement errors of 0.7±2.5 and 1.27±4.53 mmHg were obtained for the systolic and diastolic pressure, respectively. These reflect small deviations from auscultatory clinical data.
Medical subject headings
- Arterial Pressure
- Blood Pressure Monitors
- Brachial Artery