Effect of flexion and rotation on measures of coronal alignment after TKA.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 23150351.
- Also identified by DOI 10.1055/s-0032-1313756.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Attainment of a neutral mechanical axis is one of the primary goals in total knee arthroplasty (TKA). The accuracy of the procedure is evaluated by measuring the hip-knee-ankle angle (HKAA) on long leg radiographs. Rotation of the lower limb and knee flexion could possibly affect the HKAA measured on plain radiographs. In this study, a TKA was performed on a saw-bone model of the lower limb. Long leg radiographs from the femoral head to the ankle were obtained after sequentially varying the alignment of the knee model in 5-degree increments of flexion and rotation, from 0 to 20 degrees. Flexion or external rotation alone, up to 20 degrees changed the HKAA by no more than 1 degree. A combination of flexion and external rotation progressively altered the HKAA. The HKAA was altered by 5 degrees at a combination of 15- or 20-degree flexion and 20-degree external rotation. While flexion or external rotation alone had little effect on the HKAA, a combination of the two altered it substantially.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Joint
- Range of Motion, Articular
Anatomy
- knee