Usefulness of the virtual monochromatic image in dual-energy spectral CT for decreasing renal cyst pseudoenhancement: a phantom study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23169724.
- Also identified by DOI 10.2214/AJR.12.8660.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine the effect that dual-energy spectral CT has on renal cyst pseudoenhancement with a renal cyst phantom. Two renal compartments within a CT phantom were filled with 40- and 240-HU solutions of diluted contrast material, mimicking the nonenhanced phase and nephrographic phase of MDCT, respectively. A saline-filled tapering cylindric cyst model was used, simulating cysts of varying diameters (range, 0.7-3.0 cm) suspended in the renal compartments. The phantom was scanned using dual-energy CT (DECT) (80 and 140 kVp) and conventional, single-energy CT (120 kVp) with the same 64-MDCT scanner. Attenuation increase was measured for a cyst model within 40- and 240-HU dilutions. We compared the attenuation increase obtained on virtual monochromatic images from DECT with conventional, single-energy MDCT images using a paired Student t test. Pseudoenhancement occurred in the conventional single-energy CT acquisitions, with water attenuation increase of 42.44 ± 4.03 HU measured at 120 kVp. In virtual monochromatic images of DECT, we observed less pseudoenhancement with water attenuation increase of 21.51 ± 6.18 HU at 70 keV. Both acquisitions yielded a p value less than 0.0001. We achieved less pseudoenhancement of renal cysts with virtual monochromatic images of DECT compared with conventional single-energy CT. This method may be useful when an accurate differentiation between small renal cysts and solid masses is critical.
Medical subject headings
- Kidney Diseases, Cystic
- Tomography, X-Ray Computed