Nanoliter droplet viscometer with additive-free operation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23192296.
- Also identified by DOI 10.1039/c2lc41130j.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Measurement of a solution's viscosity is an important analytic technique for a variety of applications including medical diagnosis, pharmaceutical development, and industrial processing. The use of droplet-based (e.g., water-in-oil) microfluidics for viscosity measurements allows nanoliter-scale sample volumes to be used, much smaller than those either in standard macro-scale rheometers or in single-phase microfluidic viscometers. By observing the flow rate of a sample plug driven by a controlled pressure through an abrupt constriction, we achieve accurate and precise measurement of the plug viscosity without addition of labels or tracer particles. Sample plugs in our device geometry had a volume of ~30 nL, and measurements had an average error of 6.6% with an average relative standard deviation of 2.8%. We tested glycerol-based samples with viscosities as high as 101 mPa s, with the only limitation on samples being that their viscosity should be higher than that of the continuous oil phase.
Medical subject headings
- Microfluidic Analytical Techniques