Projection reconstruction magnetic particle imaging.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23193308.
- Also identified by DOI 10.1109/TMI.2012.2227121 and PMC identifier 3799838.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We acquire the first experimental 3-D tomographic images with magnetic particle imaging (MPI) using projection reconstruction methodology, which is similar to algorithms employed in X-ray computed tomography. The primary advantage of projection reconstruction methods is an order of magnitude increase in signal-to-noise ratio (SNR) due to averaging. We first derive the point spread function, resolution, number of projections required, and the SNR gain in projection reconstruction MPI. We then design and construct the first scanner capable of gathering the necessary data for nonaliased projection reconstruction and experimentally verify our mathematical predictions. We demonstrate that filtered backprojection in MPI is experimentally feasible and illustrate the SNR and resolution improvements with projection reconstruction. Finally, we show that MPI is capable of producing three dimensional imaging volumes in both phantoms and postmortem mice.
Medical subject headings
- Algorithms
- Dextrans
- Image Interpretation, Computer-Assisted
- Magnetic Resonance Imaging
- Magnetite Nanoparticles
- Whole Body Imaging