Case report: Perioperative pacemaker-mediated tachycardia in the patient with a dual chamber implantable cardioverter-defibrillator.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 23266999.
- Also identified by DOI 10.1213/ANE.0b013e3182768ce3.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Patients with cardiac implantable electronic devices are at additional risk for arrhythmias while undergoing surgical procedures. In this case report, we present a patient with a dual chamber implantable cardioverter-defibrillator who developed intraoperative pacemaker-mediated tachycardia causing significant hemodynamic instability. Management of this arrhythmia can be particularly challenging, because standard application of a magnet does not affect the pacing functions of an implantable cardioverter-defibrillator. Awareness by the anesthesiologist and timely coordination with the cardiac electrophysiology team helped to optimize care for this patient.
Medical subject headings
- Defibrillators, Implantable
- Pacemaker, Artificial
- Tachycardia