Preload sensitivity in cardiac assist devices.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 23272869.
- Also identified by DOI 10.1016/j.athoracsur.2012.07.077 and PMC identifier 3592985.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
With implantable cardiac assist devices increasingly proving their effectiveness as therapeutic options for end-stage heart failure, it is important for clinicians to understand the unique physiology of device-assisted circulation. Preload sensitivity as it relates to cardiac assist devices is derived from the Frank-Starling relationship between human ventricular filling pressures and ventricular stroke volume. In this review, we stratify the preload sensitivity of 17 implantable cardiac assist devices relative to the native heart and discuss the effect of preload sensitivity on left ventricular volume unloading, levels of cardiac support, and the future development of continuous-flow total artificial heart technology.
Medical subject headings
- Heart Failure
- Heart-Assist Devices
- Ventricular Pressure