Surgical repair of a left ventricular pseudoaneurysm is challenging. Percutaneous closure with an AMPLATZER Septal Occluder (AGA Medical Corporation, Golden Valley, MN) has evolved to be an effective alternative to conventional surgery with less complication. We report a case of severe hemolytic hyperbilirubinemia and acute renal failure after percutaneous left ventricular pseudoaneurysm endoprothesis repair, which required conventional surgical repair through a left thoracotomy to treat this complication. The rationale and results of such an off-the-label usage of AMPLATZER Septal Occluder were reexamined. Residual leakages were not rare findings in the reviewed literature. The indications for this approach should be individualized.
Medical subject headings
- Aneurysm, False
- Heart Aneurysm
- Hyperbilirubinemia
- Septal Occluder Device