Repeat LISS treatment for femoral shaft fractures due to hardware failure: a retrospective analysis of eleven cases.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 23412207.
- Also identified by DOI 10.1007/s00590-012-1075-z.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the effectiveness of a replating technique having a less-invasive stabilization system (LISS) for femoral shaft fractures due to LISS failure in adults. There were 11 patients with hardware failure of LISS for femoral shaft fractures, on an average of 50 days after the primary operation. The failed implants were removed, and the fractures were replated with a LISS following the rationale of biological osteosynthesis. Radiological fracture union and incidence of postoperative complications were employed to evaluate the effectiveness of this replating technique for femoral shaft fractures. Operative duration including removing failed hardware and replating fractures averaged 81.5 min, with an average blood loss of 330 ml. Patients had an average follow-up of 25.7 months. Radiological evaluation indicated that fracture union occurred in an average of 4.4 months in all patients. The length and alignment of the affected limb were satisfactory, and hardware failure did not recur. The replating technique with LISS for femoral shaft fractures due to hardware failure of LISS can obtain satisfactory results when the appropriate rationale of biological osteosynthesis and functional exercise is followed.
Medical subject headings
- Bone Plates
- Femoral Fractures
- Fracture Fixation, Internal
- Prosthesis Failure
Anatomy
- femur