Blood glucose measurement by multiple attenuated total reflection and infrared absorption spectroscopy.
Where this comes from
- Record sourced from PubMed, PMID 2345001.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The difficulty of measuring physiological concentrations of glucose in blood by conventional infrared absorption spectroscopy is due to the intrinsic high background absorption of water. This limitation can be largely overcome by the use of a CO2 laser as an infrared source in combination with a multiple attenuated total reflection (ATR) technique. To demonstrate the applicability of this technique, we compared in vitro measurements of glucose in blood obtained from an experimental infrared laser spectrometer with independent measurements made by a standard YSI 23A laboratory glucose analyzer. The capability of continuous measurement of blood glucose concentration is of primary importance in the future development of a glucose sensor for diabetic patients.
Medical subject headings
- Blood Glucose
- Spectrophotometry, Infrared