Gapping fields on sloping surfaces.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 2347725.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The accurate positioning of treatment fields has always been difficult, and becomes even more important when two fields are adjacent and a gap must be calculated. The standard formula (based on similar triangles) assumes a flat surface. In Hodgkin's patients the gapping between the mantle and para-aortic fields involve sloping surfaces and the patient thickness can vary as much as 5 or 6 cm between central axis and the lower edge of the mantle field. Due to the divergence of a large mantle field, using the standard formula and gapping on the skin surface can lead to an overlap of 7 to 9 mm and a region which can receive 140% of the midplane dose. A modification of the existing formula which takes into account sloping surfaces has been formulated for both source-skin-distance and isocentric treatments. In addition, problems in positioning related to mantle fields when treating the patient prone and supine have been investigated.
Medical subject headings
- Radiotherapy