Metal on metal total hip arthroplasty using modular acetabular shells.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 23489729.
- Also identified by DOI 10.1016/j.arth.2012.10.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Metal on metal bearings were reintroduced into the United States for use in total hip arthroplasty approximately a decade ago. We evaluated a consecutive series cohort of 148 patients with 169 hips who underwent THR using a metal on metal bearing surface with a modular acetabular shell that had fixation augmentation with dome screws. At 3 to 8 year follow-up, average 4.7 years, only one component was revised (for femoral loosening). 3 hips demonstrated proximal femoral osteolysis and 3 hips demonstrated pelvic osteolysis. The average follow-up Harris Hip Score was 94 (range, 57-100). Although the results have been excellent, we continue to closely monitor these hips for any long term failures, or concerns.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Metal-on-Metal Joint Prostheses
Anatomy
- hip
- pelvis