Irreversible electroporation ablation of the liver can be detected with ultrasound B-mode and elastography.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23489942.
- Also identified by DOI 10.1016/j.surg.2012.11.022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Irreversible electroporation (IRE) is a novel ablation technique that induces permanent membrane permeability and cell death. We are interested in ultrasound B-mode and elastography to monitor IRE ablation in the liver. Yorkshire pigs underwent IRE ablation of the liver and were imaged with ultrasound B-mode and elastography. Histologic evaluation of cell death by triphenyltetrazolium chloride and hematoxylin and eosin staining was performed. Elastography showed that liver ablated by IRE exhibited increased tissue stiffness with a peak strain ratio of 2.22. The IRE lesion had a discrete border without bubble artifact, and the lesion size significantly correlated with area of cell death on histology. IRE ablation was unaffected by presence of large blood vessels or bile ducts. IRE ablation led to increased tissue stiffness that was detectable by elastography and indicative of cell death. Elastography may complement B-mode ultrasonography to monitor IRE ablation of the liver.