Do clinical examination gloves provide adequate electrical insulation for safe hands-on defibrillation? II: Material integrity following exposure to defibrillation waveforms.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 23507465.
- Also identified by DOI 10.1016/j.resuscitation.2013.03.012.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Maintaining contact with the patient during defibrillator discharge has been proposed as a method for reducing no flow time but carries an associated risk of electrocution of the rescuer. This study describes an investigation to determine if typical clinical examination gloves possess the dielectric strength needed to prevent breakdown at defibrillation voltages; a factor essential to protect the rescuer. Four types of examination glove typically used in a clinical environment were tested with two types of defibrillation waveform commonly used. For each type of glove, 10 samples were tested initially using a monophasic defibrillation waveform and then, using a fresh sample of gloves, with a Biphasic waveform. For each glove the number of shocks required before electrical breakdown occurred was recorded. Kimberly Clark KC300 (nitrile), Kimberly Clark KC500 purple (nitrile), PH Medisavers GN90 (nitrile) and Bodyguards GL6622 (Vinyl) were tested using a monophasic defibrillation waveform and broke down after a median of 1, 4.5, 1 and 1 shocks respectively. The equivalent values for Biphasic defibrillator were 2, >10, 2.5 and 1 shocks. Typical clinical examination gloves do not possess the dielectric strength required to protect a rescuer from defibrillation voltages during hands-on chest compressions.
Medical subject headings
- Defibrillators
- Electric Impedance
- Gloves, Protective
- Heart Arrest
- Materials Testing