Image quality and radiation dose stratified by patient heart rate for coronary 64- and 320-MDCT angiography.

Fujimoto, Shinichiro; Matsutani, Hideyuki; Kondo, Takeshi; Sano, Tomonari; Kumamaru, Kanako; Takase, Shinichi; Rybicki, Frank J · AJR Am J Roentgenol · 2013

retrospective_cohort · Level III

Where this comes from

Abstract

OBJECTIVE. The purpose of this study is to retrospectively measure and compare estimated radiation doses between consecutive patient cohorts who underwent coronary imaging CT with 64- and 320-MDCT scanners. MATERIALS AND METHODS. Subjects without arrhythmia (n = 4475) underwent imaging with 64-MDCT (n = 770) and 320-MDCT (n = 3705) scanners and were classified into one of five subgroups according to the patient heart rate and the image acquisition strategy. For all patients, image quality was subjectively evaluated using a 3-point scale. Estimated radiation dose and image quality were compared between subjects stratified by CT scanner and by subgroups imaged with each technology. RESULTS. For patients with a heart rate of 60 beats/min or less, the estimated radiation dose was halved (3.8 ± 2.0 vs 7.6 ± 2.6 mSv) when the 320-MDCT scanner (n = 2787) replaced the 64-MDCT scanner (n = 511). For the entire cohort, image quality score was significantly better (2.9 ± 0.4 vs 2.8 ± 0.5; p < 0.0001) and the effective dose was significantly lower (4.9 ± 3.3 vs 9.9 ± 5.4 mSv; p < 0.0001) for 320-MDCT scanners, compared with 64-MDCT scanners. CONCLUSION. Wide area-detector coronary CT angiography protocols have reduced radiation dose, with image quality maintained at the same level, compared with 64-MDCT technologies.