The influence of head size on corrosion and fretting behaviour at the head-neck interface of artificial hip joints.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 23528551.
- Also identified by DOI 10.1016/j.arth.2012.10.017.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The primary goal of this study was to determine if head size affects corrosion and fretting behaviour at the head-neck taper interface of modular hip prostheses. Seventy-four implants were retrieved that featured either a 28 mm or a 36 mm head with a metal-on-polyethylene articulation. The bore of the heads and the neck of the stems were divided into eight regions each and graded by three observers for corrosion and fretting damage separately using modified criteria as reported in the literature. The 36 mm head size featured a significant difference in the corrosion head scores (p=0.022) in comparison to the 28 mm heads. This may be attributed to a greater torque acting along the taper interface due to activities of daily living.
Medical subject headings
- Hip Prosthesis
- Prosthesis Design
- Prosthesis Failure
Anatomy
- hip