Assessment of reactive synovitis in rotating-platform posterior-stabilized design: a 10-year prospective matched-pair MRI study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 23528558.
- Also identified by DOI 10.1016/j.arth.2013.01.023.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This is the first long-term (mean 11.6 years), prospective, matched-pair study (based on age, gender, BMI and UCLA scores) using MAVRIC (multi-acquisition variable-resonance image combination) magnetic resonance imaging to analyze reactive synovitis and osteolysis between rotating-platform posterior-stabilized (RP-PS), fixed-bearing metal-back (FB-MB), and all-polyethylene tibial (APT) in active patients (24 total, 8 in each group, mean age of 64 years, mean UCLA of 8.5) with identical femoral component and polyethylene. Reactive synovitis was observed in 6 RP-PS (75%), all 8 FB-MB (100%), and 6 APT (75%). There was a significant difference between the RP-PS and FB-MB knees in volumetric synovitis (P=0.023). Osteolysis with bone loss more than 4mm was seen in 3 FB-MB, 2 APT and none for RP-PS. These were not statistically significant.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Osteoarthritis, Knee
- Prosthesis Design
- Synovitis
Anatomy
- knee