Accuracy of dual-energy radiographic absorptiometry of the lumbar spine: cadaver study.

Ho, C P; Kim, R W; Schaffler, M B; Sartoris, D J · Radiology · 1990

biomechanical · Level V

Where this comes from

Abstract

Dual-energy radiographic absorptiometry (DRA) was used to measure the bone mineral content and area density of lumbar vertebrae (L2-L3) in 11 cadavers. These data were subsequently compared with measured ash content and density. Excellent correlation was obtained between bone mineral content measured with DRA and ash weight (r = .963, P less than .0001). The accuracy error in determining mineral content in lumbar vertebrae with DRA was about 9%. In addition, strong correlation was observed between bone mineral density measured with DRA and ash density (r = .881, P less than .0001).