The use of a zero-profile device compared with an anterior plate and cage in the treatment of patients with symptomatic cervical spondylosis: A preliminary clinical investigation.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 23539708.
- Also identified by DOI 10.1302/0301-620X.95B4.30992.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In a retrospective cohort study we compared the clinical outcome and complications, including dysphagia, following anterior cervical fusion for the treatment of cervical spondylosis using either a zero-profile (Zero-P; Synthes) implant or an anterior cervical plate and cage. A total of 83 patients underwent fusion using a Zero-P and 107 patients underwent fusion using a plate and cage. The mean follow-up was 18.6 months (SD 4.2) in the Zero-P group and 19.3 months (SD 4.1) in the plate and cage group. All patients in both groups had significant symptomatic and neurological improvement. There were no significant differences between the groups in the Neck Disability Index (NDI) and visual analogue scores at final follow-up. The cervical alignment improved in both groups. There was a higher incidence of dysphagia in the plate and cage group on the day after surgery and at two months post-operatively. All patients achieved fusion and no graft migration or nonunion was observed. When compared with the traditional anterior cervical plate and cage, the Zero-P implant is a safe and convenient procedure giving good results in patients with symptomatic cervical spondylosis with a reduced incidence of dysphagia post-operatively.
Medical subject headings
- Bone Plates
- Prostheses and Implants
- Spinal Fusion
- Spondylosis
Anatomy
- cervical spine