A reduced-modulus acrylic bone cement: preliminary results.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2355302.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Excessive local contact stress is implicated as an important factor in the initiation of the loosening process after total joint arthroplasties. A reduced-modulus acrylic bone cement, which decreases the bone-cement interface stresses, was developed to test this hypothesis. The formulation consists of butylmethacrylate beads, having a glass transition temperature of 27 degrees C, in a methylmethacrylate matrix. This cement, polybutylmethylmethacrylate (PBMMA), has an elastic modulus one-eighth that of standard PMMA bone cement, 0.27 vs. 2.1 GPa, at body temperature. In vivo use in a pilot study using the sheep total hip arthroplasty model shows a reduction in the rate of loosening of femoral components when compared both radiographically and mechanically with PMMA controls.
Medical subject headings
- Acrylic Resins
- Bone Cements