The reduction of implant-related errors and waste in total knee arthroplasty using a novel, computer based, e.Label and compatibility system.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 23618749.
- Also identified by DOI 10.1016/j.arth.2013.03.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Implant related errors and wasted arthroplasty implants represent a significant problem in the healthcare system. Studies estimate that an implant is wasted in two to five percent of total joint replacement procedures. After the introduction of a standardized time-out procedure, our institution still had a wasted implant in one out of every 20 arthroplasty surgeries. We then implemented a computer based, e.Label and compatibility system which standardized the implant labeling, confirmed correct size and side, and ensured compatibility within implant systems. Use of this system decreased wasted implants from 5.7% to 0.8% of total knee arthroplasty cases. This translated to an annualized cost savings of over $75,000 over our study period which could extrapolate to over $82 million per year in savings to the US healthcare system.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Hospital Information Systems
- Knee Prosthesis
- Medical Errors
- Product Labeling