The effect of fracture stability on the performance of locking plate fixation in periprosthetic femoral fractures.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 23642449.
- Also identified by DOI 10.1016/j.arth.2013.03.022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Periprosthetic femoral fracture (PFF) fixation failures are still occurring. The effect of fracture stability and loading on PFF fixation has not been investigated and this is crucial for optimum management of PFF. Models of stable and unstable PPFs were developed and used to quantify the effect of fracture stability and loading in a single locking plate fixation. Stress on the plate was higher in the unstable compared to the stable fixation. In the case of unstable fractures, it is possible for a single locking plate fixation to provide the required mechanical environment for callus formation without significant risk of plate fracture, provided partial weight bearing is followed. In cases where partial weight bearing is unlikely, additional biological fixation could be considered.
Medical subject headings
- Bone Plates
- Femoral Fractures
- Fracture Fixation, Internal
- Periprosthetic Fractures
Anatomy
- femur