The effectiveness of a near-infrared vascular imaging device to support intravenous cannulation in children with dark skin color: a cluster randomized clinical trial.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 23649104.
- Also identified by DOI 10.1213/ANE.0b013e31828e5bde.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Poor vein visibility can make IV cannulation challenging in children with dark skin color. In the operating room, we studied the effectiveness of a near-infrared vascular imaging device (VascuLuminator) to facilitate IV cannulation in children with dark skin color. In the operating room of a general hospital in Curacao, all consecutive children (0-15 years of age) requiring IV cannulation were included in a pragmatic cluster randomized clinical trial. The VascuLuminator was made available to anesthesiologists at the operating complex in randomized clusters of 1 week. Success at first attempt was 63% (27/43, 95% confidence interval [CI], 47%-77%) in the VascuLuminator group vs 51% (23 of 45 patients, 95% CI, 36%-66%) in the control group (P = 0.27). Median time to successful cannulation was 53 seconds (interquartile range: 34-154) in the VascuLuminator group and 68 seconds (interquartile range: 40-159) in the control group (P = 0.54), and hazard ratio was 1.12 (95% CI, 0.73-1.71). The VascuLuminator has limited value in improving success at first attempt of facilitating IV cannulation in children with dark skin color.
Medical subject headings
- Catheterization, Peripheral
- Spectroscopy, Near-Infrared
- Vascular Access Devices
- Veins