Failure analysis of alumina on alumina total hip arthroplasty with a layered acetabular component: minimum ten-year follow-up study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 23684661.
- Also identified by DOI 10.1016/j.arth.2013.04.027.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This prospective study reports the outcome of total hip arthroplasty (THA) performed in a consecutive series of patients using an alumina ceramic on a ceramic-layered component (Alumina-Bearing-Surface system). The cohort consisted of 270 hips in 229 patients. The study evaluated the clinical and radiological results over a mean follow-up of 11.4 years. Revision THA was performed on 58 hips, including alumina alternative failure in 50 hips, loosening in 4 hips, recurrent dislocation in 2 hips, stem neck fracture in 1 hip and hematoma in 1 hip. The survival rate was 68% with revision for any reason as the end point. The risk factors of implant failure are the preoperative range of motion of the hip joint and postoperative dislocation.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Equipment Failure Analysis
- Hip Prosthesis
Anatomy
- hip