Influence of gap balance on the sagittal movement of a specific mobile bearing floating platform design in total knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 23688853.
- Also identified by DOI 10.1016/j.arth.2013.04.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We analyzed 119 knees implanted with mobile bearing floating platform prostheses using the navigation-assisted gap balancing technique to analyze the relationship between intraoperative sagittal movement of floating platforms and soft tissue balancing. The 95 (79.8%) knees were classified into the positive rollback group (mean insert posterior rollback 5.86 ± 1.24 mm), and the remaining 24 (20.2%) into the negative rollback group. Lateral flexion gap (LFG) differed significantly between knees with positive and negative rollback (20.5 ± 1.7 mm vs 22.1 ± 1.7 mm, P = .021). Only LFG significantly influenced the occurrence of bearing sagittal movement. Sagittal translation of the insert occurred in about 80% of knees implanted with mobile bearing floating platforms in TKA, and was affected by flexion gaps, especially on the lateral side.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Osteoarthritis, Knee
- Prosthesis Design
- Range of Motion, Articular
Anatomy
- knee