Distal femoral locking plates and the posterolateral corner.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23696254.
- Also identified by DOI 10.1055/s-0033-1347360.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to evaluate the effect of minimally invasive submuscular placement of a distal femoral locking plate on the posterolateral structures of the knee. Eight fresh-frozen cadaveric knees were dissected after application of a lateral distal femoral locking plate through a minimally invasive submuscular approach. The lateral collateral ligament and popliteus tendon were identified and inspected for injury. Distances from the plate to the lateral collateral ligament and popliteus insertions were determined.Neither the lateral collateral ligament nor the popliteus tendon was disrupted by the minimally invasive submuscular application of distal femoral periarticular locking plates. The mean distances to the lateral collateral ligament and popliteus tendon insertions were 2.5 and 6.6 mm, respectively.Distal femoral locking plates can be applied in a minimally invasive manner without disrupting the posterolateral structures of the knee.
Medical subject headings
- Arthroplasty
- Bone Plates
- Knee Joint
Anatomy
- femur
- knee