Malfunction of a double-balloon carotid shunt as a result of herniation of the proximal balloon.

Eidt, J F; Kahn, M B; Barone, G W; Cook, J M; Barnes, R W · J Vasc Surg · 1990

case_report · Level V

Where this comes from

Abstract

Successful use of a shunt during carotid endarterectomy presupposes the avoidance of known shunt-related complications. We report the failure of a double-balloon intraluminal carotid shunt caused by herniation of the proximal balloon over the end of the shunt resulting in obstruction of blood flow in the shunt.