Optical coherence tomography of the suprachoroid after CyPass Micro-Stent implantation for the treatment of open-angle glaucoma.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 23743436.
- Also identified by DOI 10.1136/bjophthalmol-2012-302951.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the supraciliary space (SCS) with anterior segment optical coherence tomography (OCT) imaging after CyPass Micro-Stent implantation. The SCS was imaged with OCT after microstent implantation at 1, 6 months, and 1 year. Images were graded on a scale of 0-4 for morphological features indicative of fluid presence within, or drainage through, the SCS. 35 patients underwent ab-interno microstent implantation. Mean age was 68.6±10.2 years. Baseline mean intraocular pressure (IOP) was 21.9±6.1 mm Hg on average of 3.0 topical medications. At 1 month, the fluid space grade was ≥1 for 96% (24/25) of patients for tenting, 79% (15/19) for fluid posterior to the microstent, and 89% (8/9) for fluid surrounding the microstent. The mean (composite) score for all features was 2.5±0.99. The majority of patients maintained aqueous fluid through 12 months. OCT imaging provides adequate visualisation of the angle, the SCS and aqueous fluid drainage after implantation of a suprachoroidal microstent into the SCS.
Medical subject headings
- Glaucoma Drainage Implants
- Glaucoma, Open-Angle
- Stents
- Tomography, Optical Coherence