Excellent survivorship of the Morscher monoblock cup with a 28-mm Metasul-on-Metasul bearing at a mean of 5-year follow-up.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 23768917.
- Also identified by DOI 10.1016/j.arth.2013.05.005.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Morscher monoblock cup is designed for optimal osteointegration and holds the potential for excellent long-term survival. The Metasul-on-Metasul bearing was introduced to eliminate wear-induced periprosthetic osteolysis. This study reviews 137 Morscher cups with a 28-mm Metasul-on-Metasul bearing used in primary THA at 5.1years (3.2-6.6) after implantation. One hundred thirty-seven cups were implanted in 119 patients. Three patients (2.2%) had died of unrelated causes, and 10 patients (8.0%) were lost to follow-up. The mean Harris Hip score was 89.8 and the mean WOMAC sum score was 18.5. No cases of acetabular osteolysis were found. The 5-year cup survival rate was 98.5%. The Morscher cup with a 28-mm Metasul-on-Metasul bearing used in uncemented THA showed no acetabular osteolysis and promising survivorship at intermediate-term follow-up.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Hip Joint
- Hip Prosthesis
- Joint Diseases
Anatomy
- hip