Subsidence of SB Charité total disc replacement and the role of undersizing.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 23771503.
- Also identified by DOI 10.1007/s00586-013-2864-8 and PMC identifier 3804687.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A possible complication after total disc replacement (TDR) is subsidence, presumably caused by asymmetric implantation, implant undersizing or reduced bone quality. This study aims to quantify the degree of subsidence of an SB Charité TDR, and investigate whether undersizing is related to subsidence. A custom developed software package (Mathworks) reconstructed 3D bone-implant geometry. A threshold for subsidence was determined by comparing penetrated bone volume (PBV) and rotation angles. Inter- and intra-observer reproducibilities were calculated. Subsidence was correlated to undersizing. High inter- and intra-observer correlation coefficients were found for the method (R > 0.92). Subsidence was quantified as PBV 700 mm(3) combined with a rotation angle >7.5°. A reduced risk of subsidence was correlated to >60 and >62 % of the bony endplate covered by the TDR endplate for L4 and L5, respectively. A reproducible method to determine undersizing was developed. Thresholds were determined related to a reduced risk of subsidence.
Medical subject headings
- Intervertebral Disc
- Intervertebral Disc Degeneration
- Lumbar Vertebrae
- Prosthesis Failure
- Prosthesis Fitting
- Total Disc Replacement
Anatomy
- lumbar spine