Dosimetric impact of the interplay effect during stereotactic lung radiation therapy delivery using flattening filter-free beams and volumetric modulated arc therapy.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23773394.
- Also identified by DOI 10.1016/j.ijrobp.2013.03.038.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated the dosimetric impact of the interplay effect during RapidArc stereotactic body radiation therapy for lung tumors using flattening filter-free (FFF) beams with different dose rates. Seven tumors with motion ≤20 mm, treated with 10-MV FFF RapidArc, were analyzed. A programmable phantom with sinusoidal longitudinal motion (30-mm diameter "tumor" insert; period = 5 s; individualized amplitude from planning 4-dimensional computed tomography) was used for dynamic dose measurements. Measurements were made with GafChromic EBT III films. Plans delivered the prescribed dose to 95% of the planning target volume, created by a 5-mm expansion of the internal target volume. They comprised 2 arcs and maximum dose rates of 400 and 2400 MU/min. For 2400 MU/min plans, measurements were repeated at 3 different initial breathing phases to model interplay over 2 to 3 fractions. For 3 cases, 2 extra plans were created using 1 full rotational arc (with contralateral lung avoidance sector) and 1 partial arc of 224° to 244°. Dynamic and convolved static measurements were compared by use of gamma analysis of 3% dose difference and 1 mm distance-to-agreement. For 2-arc 2400 MU/min plans, maximum dose deviation of 9.4% was found in a single arc; 7.4% for 2 arcs (single fraction) and <5% and 3% when measurements made at 2 and 3 different initial breathing phases were combined, simulating 2 or 3 fractions. For all 7 cases, >99% of the area within the region of interest passed the gamma criteria when all 3 measurements with different initial phases were combined. Single-fraction single-arc plans showed higher dose deviations, which diminished when dose distributions were summed over 2 fractions. All 400 MU/min plans showed good agreement in a single fraction measurement. Under phantom conditions, single-arc and single-fraction 2400 MU/min FFF RapidArc lung stereotactic body radiation therapy is susceptible to interplay. Two arcs and ≥2 fractions reduced the effect to a level that appeared unlikely to be clinically significant.
Medical subject headings
- Lung Neoplasms
- Movement
- Phantoms, Imaging
- Radiosurgery
- Radiotherapy, Intensity-Modulated