An optical model to describe image contrast with bifocal intraocular lenses.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 2378382.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We attempted to quantify the decrease in contrast associated with the concentric-style bifocal intraocular lens by using a model eye and bifocal intraocular lens, with a model unifocal intraocular lens as a control. When imaging near objects, pupils smaller than 2.5 mm produced image contrast of 100% of the control; larger pupils degraded contrast to 25% for a 6-mm pupil. For distant objects, pupils smaller than 2.5 mm produced image contrast of 70% to 95% because of the pinhole effect; larger pupils engaged the distance portion of the intraocular lens and maintain 80% contrast on average. Thus, the bifocal intraocular lens produced image contrast greater than 70% in all situations tested, except when imaging near objects with a pupil larger than 3.5 mm. The data suggest that image contrast is highly dependent on pupil size and object distance.
Medical subject headings
- Contrast Sensitivity
- Lenses, Intraocular
- Models, Biological