Range of motion in a modular femoral stem system with a variety of neck options.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 23886407.
- Also identified by DOI 10.1016/j.arth.2012.10.011.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Modular femoral stem systems decouple leg length, offset, and version. The hip ROM and type of impingement for 162 femoral head/neck combinations were measured at four extreme hip positions in a Sawbones pelvis and femur to identify constructs that lead to early impingement. Hip ROM increased in all positions with increasing head size and neck length. We identified a new type of impingement created by the build-up of the proximal femoral stem: femoral stem on acetabular liner impingement. Seventy percent of neutral neck options achieved our definition of acceptable ROM. In general, when utilizing a modular femoral stem, surgeons can minimize impingement by choosing the longest femoral neck that does not over-lengthen the limb, using the largest femoral head accommodated by the cup, and avoiding neck version unless the cup or stem is malaligned.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femur
- Hip Prosthesis
Anatomy
- femur
- hip