Towards BirthAlert--A Clinical Device Intended for Early Preterm Birth Detection.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23893706.
- Also identified by DOI 10.1109/TBME.2013.2272601 and PMC identifier 4605421.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Preterm birth causes 1 million infant deaths worldwide every year, making it the leading cause of infant mortality. Existing diagnostic tests such as transvaginal ultrasound or fetal fibronectin either cannot determine if preterm birth will occur in the future or can only predict the occurrence once cervical shortening has begun, at which point it is too late to reverse the accelerated parturition process. Using iterative and rapid prototyping techniques, we have developed an intravaginal proof-of-concept device that measures both cervical bioimpedance and cervical fluorescence to characterize microstructural changes in a pregnant woman's cervix in hopes of detecting preterm birth before macroscopic changes manifest in the tissue. If successful, such an early alert during this "silent phase" of the preterm birth syndrome may open a new window of opportunity for interventions that may reverse and avoid preterm birth altogether.
Medical subject headings
- Cervical Ripening
- Cervix Uteri
- Optical Imaging
- Premature Birth