The effect of helium on endotracheal tube flammability.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 2393138.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effect on endotracheal tube flammability of helium (He)-diluted oxygen and He-diluted nitrous oxide was determined with the oxidant O2 (He) and oxidant N2O (He) indices of flammability. These values then were compared with the corresponding values for nitrogen dilution, i.e., oxidant O2 (N2) and oxidant N2O (N2) indices of flammability. Four different types of endotracheal tubes were studied: polyvinylchloride (PVC), red rubber (RR), silicone (Si), and the Xomed (Xo) Lasershield tube. The oxidant O2 (He) indices are: PVC 0.274 +/- 0.0055, RR 0.194 +/- 0.0089, Si 0.194 +/- 0.0055, and Xo 0.256 +/- 0.0055. The oxidant N2O (He) indices are: PVC 0.526 +/- 0.0055, RR 0.434 +/- 0.0114, Si 0.416 +/- 0.0055, and Xo 0.456 +/- 0.0154. The oxidant O2 (N2) indices are: PVC 0.254 +/- 0.0055, RR 0.182 +/- 0.0045, Si 0.200 +/- 0, and Xo 0.230 +/- 0. The oxidant N2O (N2) indices are: PVC 0.472 +/- 0.0084, RR 0.356 +/- 0.0055, Si 0.392 +/- 0.0045, and Xo 0.444 +/- 0.0114. These differences, although statistically significant, are modest and demonstrate only a small, and probably clinically not significant, preference of He over N2 in decreasing flammability.
Medical subject headings
- Fires
- Helium
- Intubation, Intratracheal