Metal ion trend may be more predictive for malfunctioning metal-on-metal implants than a single measurement.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 23934907.
- Also identified by DOI 10.5301/hipint.5000066.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Forty-eight unilateral hip resurfacing arthroplasty patients were evaluated for cobalt and chromium levels. The metal ion trend of 42 well-functioning patients was compared with six sub-optimal functioning patients. Median metal ion levels were significantly higher for the sub-optimal group. For the well-functioning implants, the percentage of patients with increasing cobalt/chromium levels between two consecutive time-intervals ('risers') gradually decreased from 90/86% (0-3 months) to 22/22% (24-36 months). The percentage of patients with increasing metal ion levels was higher in the sub-optimal group. The median absolute increase of this 'risers' subgroup was significantly lower for the well-functioning group at 12-24 months. Sub-optimal functioning MoM implants have a different metal ion trend than well-functioning implants, a higher chance of 'risers' and a larger absolute increase in time.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Chromium
- Cobalt
- Hip Prosthesis
- Metal-on-Metal Joint Prostheses
- Osteoarthritis, Hip
Anatomy
- hip