Influence of pupil size on measurements made with the Lens Opacity Meter 701.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 2393643.
- Also identified by PMC identifier 1042199.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Lens Opacity Meter 701 (Interzeag, Switzerland) is a recently developed instrument for assessing cataract. It measures the degree of scatter of a red light beam by the lens. Fifty patients prior to cataract surgery had measurements made before and after dilatation of the pupil. Inconsistent results were found at pupil diameters of less than 4 mm. It is postulated that this was due to absorption of scattered light by the iris.