Filling the void: a coalescent numerical and experimental technique to determine aortic stent graft mechanics.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 23953501.
- Also identified by DOI 10.1016/j.jbiomech.2013.07.010.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The presented study details a combined experimental and computational method to assess and compare the mechanical behavior of the main body of 4 different stent graft designs. The mechanical response to a flat plate compression and radial crimping of the devices is derived and related to geometrical and material features of different stent designs. The finite element modeling procedure is used to complement the experimental results and conduct a solution sensitivity study. Finite element evaluations of the mechanical behavior match well with experimental findings and are used as a quantitative basis to discuss design characteristics of the different devices.
Medical subject headings
- Prosthesis Design
- Stents