Robotic tissue tracking for beating heart mitral valve surgery.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23973122.
- Also identified by DOI 10.1016/j.media.2010.06.007 and PMC identifier 3795906.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The rapid motion of the heart presents a significant challenge to the surgeon during intracardiac beating heart procedures. We present a 3D ultrasound-guided motion compensation system that assists the surgeon by synchronizing instrument motion with the heart. The system utilizes the fact that certain intracardiac structures, like the mitral valve annulus, have trajectories that are largely constrained to translation along one axis. This allows the development of a real-time 3D ultrasound tissue tracker that we integrate with a 1 degree-of-freedom (DOF) actuated surgical instrument and predictive filter to devise a motion tracking system adapted to mitral valve annuloplasty. In vivo experiments demonstrate that the system provides highly accurate tracking (1.0 mm error) with 70% less error than manual tracking attempts.