Fast Count-Regulated OSEM Reconstruction With Adaptive Resolution Recovery.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 23996543.
- Also identified by DOI 10.1109/TMI.2013.2279851.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Ordered subsets expectation maximization (OSEM) is widely used to accelerate tomographic reconstruction. Speed-up of OSEM over maximum likelihood expectation maximization (MLEM) is close to the number of subsets (NS). Recently we significantly increased the speed-up achievable with OSEM by specific subset choice (pixel-based OSEM). However, a high NS can cause undesirable noise levels, quantitative inaccuracy or even disappearance of lesions in low-activity image regions, while a low NS leads to prohibitively long reconstructions or unrecovered details in highly active regions. Here, we introduce count-regulated OSEM (CROSEM) which locally adapts the effective NS based on the estimated amount of detected photons originating from individual voxels. CROSEM was tested using multi-pinhole SPECT simulations and in vivo imaging. With the maximum NS set to 128, CROSEM attained acceleration factors close to 128 in high-activity regions and kept quantitative accuracy in low-activity regions close to that of MLEM. At equal cold-lesion contrast in high-activity regions, CROSEM exhibited lower noise than MLEM in low-activity regions. CROSEM is a fast and stable alternative to OSEM, preventing excessive image noise and quantitative errors in low-activity regions while achieving high-resolution recovery in structures with high activity uptake.