Long-term outcome of a metal-on-polyethylene cementless hip resurfacing.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 24090660.
- Also identified by DOI 10.1016/j.arth.2013.08.016.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Due to the well-documented problems surrounding metal-on-metal bearings, the use of hip resurfacing has declined. Since the potential benefits of hip resurfacing remain desirable, it may be beneficial to investigate the long-term outcome of hip resurfacings using metal-on-polyethylene in the 1980's. We report the long-term survivorship and modes of failure of a cementless metal-on-polyethylene resurfacing (n = 178) with different porous ingrowth surfaces. While acetabular loosening was absent, a high incidence of femoral failures (femoral loosening = 18.1%, osteolytic neck fracture = 21%) occurred despite using the same ingrowth surface for both components. Ongoing developments using the lessons learned from these previous generation components and utilizing modern low wear materials, e.g., cross-linked polyethylene, may lead to improved implants for future hip resurfacings.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Joint
- Hip Prosthesis
- Joint Diseases
Anatomy
- hip