Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 24090661.
- Also identified by DOI 10.1016/j.arth.2013.08.026.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Corrosion at the modular neck-stem taper junction has become an increasingly important topic as several reports have identified this couple as a possible source for early failure with findings similar to failed metal-on-metal hip arthroplasties. Recently, two different modular stem systems from a single manufacturer were voluntarily recalled due to concerns of failure of the modular taper junction. We discuss how to approach the diagnosis and management of patients with these particular stem systems. We further reviewed the literature to evaluate whether this is a manufacturer-specific defect or indicative of a broader trend. Recent studies appear to implicate the basic design of the neck-stem taper junction, rather than a single manufacturer, which is at high risk for fretting and corrosion.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Joint Diseases
- Medical Device Recalls
- Prosthesis Failure
Anatomy
- hip