Three-dimensional ultrasound improves the accuracy of diameter measurement of the residual sac in EVAR patients.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 24091092.
- Also identified by DOI 10.1016/j.ejvs.2013.09.012.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Discrepancy between maximum diameters obtained with two-dimensional ultrasound and computed tomography (CT) after endovascular aneurysm repair (EVAR) is well known. The maximal diameter is ideally measured perpendicular to the centerline, a methodology so far only feasible with three-dimensional (3D) CT and magnetic resonance angiography (MRA). We aimed to investigate the agreement between 3D ultrasound and 3D CT and to determine reproducibility measures. Prospective study comparing 3D ultrasound with 3D CT in 124 consecutive patients seen 3 or 12 month after EVAR. Replacing 2D with 3D ultrasound, the mean difference was improved from 6.0 mm to -1.3 mm (p < .001), and the range of variability was reduced from 9.4 mm to 6.6 mm (p = .009) using 3D CT as the gold standard. The mean difference between 3D ultrasound and 3D CT maximum diameter of the residual sac was -1.3 mm with upper and lower limits of agreement of 5.2 mm and -7.9 mm, respectively. Reproducibility measures of 3D ultrasound were ± 4 mm. 3D ultrasound correlate significantly better to 3D CT than the currently used 2D ultrasound method when assessing maximum diameter of the residual sac after EVAR, and reproducibility measures were within clinical acceptable values.
Medical subject headings
- Aortic Aneurysm, Abdominal
- Blood Vessel Prosthesis Implantation
- Endovascular Procedures
- Image Interpretation, Computer-Assisted
- Imaging, Three-Dimensional
- Ultrasonography, Interventional